FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 3011836 · Received March 19, 2013

Report

Report Number
3008382007-2013-05609
Event Type
Malfunction
Date Received
March 19, 2013
Report Date
February 22, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K073231
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

(B)(4): THE TEST STRIP HAS PASSED TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(4) 2013 THE LAY USER/PATIENT IN USA CONTACTED LIFESCAN TO REPORT THE ONE TOUCH ULTRALINK METER WAS GIVING INACCURATELY LOW READINGS WITH THE CONTROL SOLUTION. THE PATIENT OBTAINED THE CONTROL SOLUTION READING OF 119 MG/DL ON THE REPORTED METER. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED AND THE RESULTS FELL OUT OF RANGE FOR THE LOT OF TEST STRIPS IN USE, THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114331 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3379527

Patients

Seq Age Sex Outcome Treatment
1