ACCU-CHEK SPIRIT CARTRIDGE
Report
- Report Number
- 2183996-2013-00410
- Event Type
- Malfunction
- Date Received
- March 19, 2013
- Date of Event
- March 11, 2013
- Report Date
- August 4, 2025
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
- Health Professional
- N
Narratives
THE COMPLAINT DESCRIBING A LEAKAGE CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATIONS. ONE USED AND TEN UNUSED CARTRIDGES WERE VISUALLY, LEAK AND GLIDE FORCE TESTED. THE CARTRIDGES PASSED THE VISUAL TEST AND THE LEAKAGE TEST AT DIFFERENT POSITIONS OF THE PLUNGER ROD. THE GLIDE FORCE MEASURED IS IN ACCORDANCE WITH PRODUCT SPECIFICATIONS.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
PATIENT REPORTED INSULIN LEAKED FROM THE CARTRIDGE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED, AND SHE DID NOT REQUIRE TREATMENT FORM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE CARTRIDGE WAS REQUESTED FOR EVALUATION.