FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT ADAPTER

MDR report key: 3011780 · Received March 19, 2013

Report

Report Number
2183996-2013-00409
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
March 9, 2013
Report Date
July 24, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE COMPLAINT CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATIONS REGARDING THE CUSTOMERS ALLEGATION. THE RECEIVED ADAPTER IS AN ACCU- CHECK SPIRIT COMBO ADAPTER. IT PASSED THE OPTICAL INVESTIGATION AND COMPLIES WITH THE PRODUCT SPECIFICATIONS. ONE USED INFUSION SET WAS RECEIVED FOR EVALUATION. A FREE FLOW AND TIGHTNESS TEST ON THE CONCERNED SAMPLE WAS PERFORMED AND NO IRREGULARITIES WERE FOUND. A VISUAL INSPECTION ON THE RETENTION SAMPLE AND AN ADDITIONAL FREE FLOW AND TIGHTNESS TEST ON ONE RETENTION SAMPLE WERE PERFORMED, AND NO DEVIATIONS WERE FOUND.

Description of Event or Problem · 1

PATIENT REPORTED HE CHANGED THE ADAPTER ON (B)(6) 2013 AND EXPERIENCED HYPERGLYCEMIA THAT EVENING (VALUE NOT PROVIDED). HE CHANGED THE INFUSION SET AND DELIVERED INSULIN VIA THE INFUSION DEVICE TO TREAT HYPERGLYCEMIA. ON (B)(6) 2013, HE NOTICED THE ADAPTER CAUSED INSULIN TO LEAK. HE CHANGED THE ADAPTER AGAIN, AND THIS RESOLVED THE ISSUE. HE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE ADAPTER WAS REQUESTED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113435 ACCU-CHEK ® SPIRIT ADAPTER INSULIN INFUSION PUMP ACCESSORY LZG ROCHE DIABETES CARE AG 00700006865 ASKU

Patients

Seq Age Sex Outcome Treatment
1 038 YR