FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3011623 · Received March 19, 2013

Report

Report Number
2032227-2013-01080
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
March 5, 2013
Report Date
March 5, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CONSTANT ALARM NOTED DURING REWIND DUE TO FAULTY MOTHER BOARD. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO ALARM. BROKEN RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW AND MISSING END CAP STICKER NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED. TROUBLESHOOTING WAS PERFORMED, AND THE DEVICE FAILED THE DISPLACEMENT TEST. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114139 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAL A000114715

Patients

Seq Age Sex Outcome Treatment
1 45 YR