INTERSTIM
Report
- Report Number
- 3004209178-2013-03938
- Event Type
- Malfunction
- Date Received
- March 19, 2013
- Report Date
- February 28, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- EZW
- PMA / PMN Number
- P970004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PRODUCT ID 3889-28, LOT # V004103, IMPLANTED: (B)(6) 2006, PRODUCT TYPE LEAD; PRODUCT ID 3095-10, SERIAL # (B)(4), IMPLANTED: (B)(6) 2010, PRODUCT TYPE EXTENSION; PRODUCT ID 3037, SERIAL # (B)(4), PRODUCT TYPE PROGRAMMER, PATIENT. (B)(4).
(B)(4).
IT WAS REPORTED THAT THE PATIENT WAS HAVING DISCOMFORT IN THE POCKET AREA, "HER RIGHT ILIAC CREST AREA", AND THERE WAS A POWER ON RESET (POR) CONDITION FOLLOWING ELECTROMAGNETIC INTERFERENCE (ERI). THE CALLER (HEALTHCARE PROVIDER, HCP) ASSESSED THE POCKET AND BELIEVED SHE FELT A LEAD WIRE SUPERFICIALLY. THE HCP STATED THAT THE PAIN WAS NOT DEVICE RELATED AND IT "APPEARED TO BE THE VALVE BETWEEN THE LARGE AND SMALL INTESTINE." IT WAS STATED THAT THE PATIENT'S DEVICE WAS OFF AT THE TIME OF REPORT. IT WAS ALSO STATED THAT THE PATIENT HAD A STROKE IN (B)(6) 2012 THEN HAD AN MRI. REPORTEDLY, THE FACILITY WAS UNAWARE THE PATIENT HAD AN IMPLANTED DEVICE. FOLLOWING THIS STROKE WAS WHEN THE POR CONDITION OCCURRED AS WELL AS THE RETURN OF SYMPTOMS. THE HCP SUSPECTED THE MRI HAD SHIFTED THE INS/LEAD/EXTENSION. THE HCP ASKED THE PATIENT TO TURN ON HER DEVICE TO PROGRAM 1 BUT TURN DOWN TO 1 VOLT FROM 1.6 VOLTS PRIOR TO TURNING ON. IT WAS NOTED THAT THE PATIENT THEN FELT OVERSTIMULATION AND PRESSED THE EMERGENCY OFF BUTTON. THE PATIENT TURNED THE STIMULATION DOWN TO 0 VOLTS AND TURNED OFF THE DEVICE UNTIL ANOTHER HCP COULD ASSIST.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 114341 | INTERSTIM | STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE | EZW | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | 3023 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |