FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3011444 · Received March 19, 2013

Report

Report Number
2531779-2013-02833
Event Type
Malfunction
Date Received
March 19, 2013
Report Date
February 19, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, A REVIEW OF THE PUMP HISTORY SHOWED THE PUMP HAD REBOOTED. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. THERE WAS NO DAMAGE FOUND TO THE BATTERY COMPARTMENT. A TEST CAP WAS ABLE TO ATTACH SECURELY TO THE PUMP AND THE PUMP POWERED ON WITH A BLANK DISPLAY. A TEST DISPLAY WAS INSERTED FOR TESTING AND THE DISPLAY REMAINED BLANK. EVALUATION REVEALED THAT THE BOLUS BUTTON WAS MISALIGNED. A LEAK TEST WAS PERFORMED AND A LEAK WAS FOUND IN THE BOLUS BUTTON. THE PUMP WAS OPENED AND MOISTURE DAMAGE WAS FOUND ON THE PRINTED CIRCUIT BOARD.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 INDICATING THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 250MG/DL AFTER THE PUMP LOST POWER AFTER THE PATIENT WAS SWIMMING. THE PATIENT WAS REPORTEDLY SWIMMING AND WATER GOT INTO THE PUMP. THE PATIENT WAS GOING TO DO A MEAL BOLUS AND NOTICED THAT THE PUMP WAS POWERED OFF. THE PATIENT WAS ABLE TO TREAT THEIR BLOOD GLUCOSE WITH ANOTHER INSULIN DELIVERY METHOD. THIS REPORT IS BEING MADE BASED ON THE INDICATION THAT THE PUMP POWERED OFF WITHOUT USER INTERVENTION. THE BLOOD GLUCOSE EXCURSION DOES NOT CONSTITUTE AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114259 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 41 YR