ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-02833
- Event Type
- Malfunction
- Date Received
- March 19, 2013
- Report Date
- February 19, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, A REVIEW OF THE PUMP HISTORY SHOWED THE PUMP HAD REBOOTED. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. THERE WAS NO DAMAGE FOUND TO THE BATTERY COMPARTMENT. A TEST CAP WAS ABLE TO ATTACH SECURELY TO THE PUMP AND THE PUMP POWERED ON WITH A BLANK DISPLAY. A TEST DISPLAY WAS INSERTED FOR TESTING AND THE DISPLAY REMAINED BLANK. EVALUATION REVEALED THAT THE BOLUS BUTTON WAS MISALIGNED. A LEAK TEST WAS PERFORMED AND A LEAK WAS FOUND IN THE BOLUS BUTTON. THE PUMP WAS OPENED AND MOISTURE DAMAGE WAS FOUND ON THE PRINTED CIRCUIT BOARD.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 INDICATING THAT THE PATIENT'S BLOOD GLUCOSE LEVEL ROSE TO 250MG/DL AFTER THE PUMP LOST POWER AFTER THE PATIENT WAS SWIMMING. THE PATIENT WAS REPORTEDLY SWIMMING AND WATER GOT INTO THE PUMP. THE PATIENT WAS GOING TO DO A MEAL BOLUS AND NOTICED THAT THE PUMP WAS POWERED OFF. THE PATIENT WAS ABLE TO TREAT THEIR BLOOD GLUCOSE WITH ANOTHER INSULIN DELIVERY METHOD. THIS REPORT IS BEING MADE BASED ON THE INDICATION THAT THE PUMP POWERED OFF WITHOUT USER INTERVENTION. THE BLOOD GLUCOSE EXCURSION DOES NOT CONSTITUTE AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 114259 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |