FDA Adverse Event Injury Summary report: N

LCS COMPLETE FEM CEM L STD

MDR report key: 3011353 · Received March 19, 2013

Report

Report Number
1818910-2013-14021
Event Type
Injury
Date Received
March 19, 2013
Date of Event
March 6, 2013
Report Date
April 4, 2013
Manufacturer
9616671 DEPUY IRELAND
Product Code
NJL
PMA / PMN Number
PP830055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

VISUAL EXAMINATION OF THE RETURNED BILATERAL DEVICES FINDS THE TIBIAL INSERTS DISPLAY EXCESSIVE PITTING AND SCRATCHES ON BOTH THE ARTICULATING AND NON-ARTICULATING SURFACES. MATCHING WEAR IS FOUND ON BOTH FEMORAL AND TIBIAL COMPONENT ARTICULATING SURFACES INDICATING THE PRESENCE OF THIRD BODY PARTICULATES, MOST LIKELY DUE TO BONE CEMENT. IT WAS NOT INDICATED IF OR IF NOT THE PATIENT HAD A TRAUMA PRIOR TO REVISION. REVIEW OF PROVIDED PATIENT X-RAYS FINDS THE FEMORAL COMPONENTS HAVE SLIGHT GAP ON THE ANTERIOR FLANGES. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT AND LOT CODE COMBINATION SINCE ITS RELEASE TO DISTRIBUTION. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. PRODUCT PROBLEM HAS NOT BEEN IDENTIFIED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THIS INVESTIGATION CLOSED. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION THAT CHANGES THIS CONCLUSION BE RECEIVED, THE INVESTIGATION WILL BE REOPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

REVISION WAS DONE ON BOTH KNEES. DURING THE SURGERY, IT WAS NOTED THAT SYNOVIALIS WAS BLACKENED AND METALLOSIS EXISTED IN BOTH KNEES.

Description of Event or Problem · 1

UPDATE 4/4/13 PRODUCT/LOT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113268 LCS COMPLETE FEM CEM L STD FEMORAL NJL 9616671 DEPUY IRELAND 3151009

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention