FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3011336 · Received March 19, 2013

Report

Report Number
1416980-2013-06586
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
February 22, 2013
Report Date
February 22, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED A FLOGARD INFUSION PUMP WITH AN F-38 ALARM, WHICH WAS CONFIRMED BY TECHNICAL SERVICES. THE CAUSE WAS DETERMINED TO BE DAMAGED LEFT AND RIGHT TUBE MISLOADING SENSORS. THE TUBE MISLOADING SENSORS WERE REPLACED IN ORDER TO RESOLVE THE PROBLEM. ADDITIONAL INFORMATION: BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE FACILITY REPORTED A FLOGARD INFUSION PUMP THAT PRESENTED AN "F38 ALARM". IT IS UNKNOWN IF THIS EVENT OCCURRED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT/USER INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114000 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1