PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2013-01573
- Event Type
- Injury
- Date Received
- March 19, 2013
- Date of Event
- February 22, 2013
- Report Date
- February 25, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4):DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION.THE OTHER PERCLOSE PROGLIDE DEVICE REFERENCED IN B5 IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER.THE DEVICE WAS REPORTED TO BE DISCARDED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR INVESTIGATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENT IN THE COMPLAINT-HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF AN UNSPECIFIED VESSEL WAS ATTEMPTED USING A PROGLIDE AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, THE SUTURES WERE CUT DURING KNOT FORMATION. A SECOND PROGLIDE WAS USED WITH THE SAME RESULTS. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
SUBSEQUENT TO THE INITIAL MEDWATCH REPORT, THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE, WITH A 6FR SHEATH, AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, THE SUTURES WERE CUT DURING KNOT FORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 113917 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 20213J1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | HEPARIN| SHEATH: 6F |