FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3011192 · Received March 19, 2013

Report

Report Number
1416980-2013-06582
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
February 22, 2013
Report Date
February 22, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CUSTOMER REPORTED PROBLEM OF A FLOGARD INFUSION PUMP WITH AN F94 ALARM WAS CONFIRMED. SERVICE DETERMINED THAT THE CAUSE WAS DUE TO DAMAGED MAIN BATTERIES. THE MAIN BATTERIES WERE REPLACED TO RESOLVE THE ISSUE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE FACILITY REPORTED A FLOGARD INFUSION PUMP THAT PRESENTED AN "F94 ALARM". IT IS UNKNOWN IF THIS EVENT OCCURRED DURING PATIENT USE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT/USER INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114515 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1