FDA Adverse Event Injury Summary report: N

SORBAFIX ENHANCED

MDR report key: 3011040 · Received March 12, 2013

Report

Report Number
1213643-2013-00096
Event Type
Injury
Date Received
March 12, 2013
Date of Event
February 13, 2013
Report Date
February 19, 2013
Manufacturer
DAVOL INC., SUB. C.R. BARD, INC.
Product Code
GDW
PMA / PMN Number
K111153
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A TOTAL OF FIVE DEVICES AND ONE SEPRAMESH WERE RETURNED FOR EVAL. THE SEPRAMESH WAS RECEIVED WITH 52 TACKS DEPLOYED INTO THE MESH AND SIX OUT OF THE 52 FASTENERS WERE NOTED TO BE PROUD. IT CANNOT BE DETERMINED WHICH FASTENERS IN THE MESH WERE DEPLOYED FROM WHICH SORBAFIX DEVICE. THE THIRD DEVICE WAS RECEIVED WITH A TACK PROTRUDING FROM THE OUTER CANNULA. A TACK REMAINED PROTRUDING AFTER TWO ACTUATIONS OF THE DEVICE. THE DEVICE HANDLE COVER WAS REMOVED TO EXPOSE THE GEAR TRAIN AND INNER COMPONENTS. AN EVAL OF THE INNER COMPONENTS FOUND THE CLUTCH INPUT GEAR WAS OUT OF SYNC AND DAMAGED. TWENTY FASTENERS REMAINED WITHIN THE DEVICE. BASED ON THE AVAILABLE INFO AND THE PRODUCT EVAL, WE ARE UNABLE TO DETERMINE A DEFINITIVE ROOT CAUSE AT THIS TIME. IT IS POSSIBLE THAT EXCESSIVE RESISTANCE CAUSED THE INNER COMPONENT DAMAGE DUE TO THE FORCES APPLIED TO THE DEVICE DURING USE. A LOT NUMBER WAS NOT PROVIDED, THEREFORE, A MFG REVIEW COULD NOT BE PERFORMED. THE IFU CAUTIONS TO AVOID EXCESSIVE TRIGGER FORCE AS THIS MAY DAMAGE THE DEVICE. SEE MDR 1213643-2013-00089 FOR INFO RELATED TO THE FIRST SORBAFIX USED ON (B)(6) 2013. SEE MDR 1213643-2013-00095 FOR INFO RELATED TO THE SECOND SORBAFIX USED ON (B)(6) 2013. SEE MDR 1213643-2013-00097 FOR INFO RELATED TO THE FOURTH SORBAFIX USED ON (B)(6) 2013. SEE MDR 1213643-2013-00098 FOR INFO RELATED TO THE FIFTH SORBAFIX USED ON (B)(6) 2013.

Description of Event or Problem · 1

INFO AS PROVIDED TO DAVOL: DURING A LAPAROSCOPIC HERNIA REPAIR THE SORBAFIX DEVICE WAS DEPLOYED FOUR TIMES, AFTER WHICH THE DEVICE THEN WOULD STICK AND WOULD NOT DEPLOY FASTENERS. A TOTAL OF FIVE SORBAFIX DEVICES WERE USED, EACH WAS REPORTED AS HAVING THE SAME ISSUES WITH JAMMING AND NOT DEPLOYING FASTENERS. AFTER SEVERAL ATTEMPTS TO FIXATE THE MESH (BARD SEPRAMESH,) THE MESH WAS REMOVED. THE SURGEON CHOSE TO CONVERT FROM A LAPAROSCOPIC TO AN OPEN PROCEDURE TO FACILITATE COMPLETION OF THE CASE. THERE WAS NO REPORTED ADVERSE OUTCOME TO THE PT, ASIDE FROM HAVING TO CONVERT FROM A LAPAROSCOPIC TO AN OPEN PROCEDURE. DEVICE: 3.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104470 SORBAFIX ENHANCED GDW DAVOL INC., SUB. C.R. BARD, INC. NA NI

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention