AXON PATIENT MODULE
Report
- Report Number
- 1045254-2013-00267
- Event Type
- Malfunction
- Date Received
- March 14, 2013
- Report Date
- August 3, 2012
- Manufacturer
- MEDTRONIC XOMED, INC.
- Product Code
- GWF
- PMA / PMN Number
- K061639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
EVAL FOUND A DEFECTIVE CONNECTOR BOARD. THE MFG DATE IS UNOBTAINABLE. RECORDS PRIOR TO SEPTEMBER 2011 REMAIN WITH THE ORIGINAL DEVICE MANUFACTURE. THE NIM-ECLIPSE SYSTEM NEUROVASCULAR WORKSTATION IS INTENDED FOR USE TO MONITOR SENSORY AND MOTOR PATHWAYS. IF THE SYSTEM FAILS TO PERFORM AS INTENDED, (ESPECIALLY TO EFFECT OR DETECT ELECTROMYOGRAPHIC (EMG) ACTIVITY) IT PRESENTS THE POTENTIAL FOR TEMPORARY OR PERMANENT DAMAGE TO THE NERVE THAT IS PRESENT. THE ECLIPSE PATIENT MODULE SERVES AS AN INTERFACE BETWEEN THE PATIENT AND THE ECLC/CONTROLLER. IT IS DESIGNED TO DELIVER STIMULUS AND RECORD THE READINGS VIA ELECTRODES. WHEN INFO SUGGESTS THAT THE NIM ECLIPSE EQUIPMENT HAD THE POTENTIAL TO NOT STIMULATE TO AFFECT ELECTROMYOGRAPHY (EMG), OR TO DETECT EMG, IT IS ASSUMED THAT THE FAILURE WAS DEVICE-RELATED AND MAY HAVE GONE UNDETECTED. IN THIS CASE, THERE IS NO INFO TO SUGGEST AN EVENT COULD NOT BE INTERPRETED FALSELY- THE REPORT IS INCONCLUSIVE AS TO THE CAUSE OF FIELD OBSERVATION. THEREFORE, WITHOUT INFO TO REASONABLY SUGGEST SERIOUS INJURY, OR THAT MEDICAL INTERVENTION WAS REQUIRED TO PRECLUDE SERIOUS INJURY, WE ARE FILING THIS REPORT AS A PRODUCT PROBLEM. THE NIM-ECLIPSE SYSTEM DOES NOT PREVENT THE SURGICAL SEVERING OF NERVES. IF MONITORING IS COMPROMISED, THE SURGICAL PRACTITIONER MUST RELY ON ALTERNATE METHODS, OR SURGICAL SKILLS, EXPERIENCE, AND ANATOMICAL KNOWLEDGE TO PREVENT DAMAGE TO NERVES. THIS PRODUCT WAS BEING USED FOR TREATMENT, NOT DIAGNOSIS.
THE MFR HAS REFINED THE CRITERIA FOR MAKING MDR DECISIONS. UPON A RETROSPECTIVE REVIEW, THE FOLLOWING EVENT IS NOW BELIEVED TO BE REPORTABLE: A CUSTOMER RETURNED A PATIENT INTERFACE MODULE FOR SERVICE/REPAIR STATING "NON-SPECIFIC BROKEN." THERE WAS NO SUGGESTION OF PATIENT INJURY. TESTING/REPAIR FOUND A DEFECTIVE CONNECTOR BOARD. THE AVAILABLE INFO INDICATES THE PATIENT MODULE HAD A NON-SPECIFIC MALFUNCTION, WHICH SUGGESTS THAT IT MAY NOT HAVE BEEN STIMULATING/DELIVERING CURRENT, AND THE USER WAS NOT ALERTED BY A SYSTEM ALARM, WARNING SCREEN OR ERROR MESSAGE. EVENT CONTEXT WAS NOT GIVEN. IF THE SYSTEM IS NOT STIMULATING/DELIVERING CURRENT AND THE USER HAS NOT BEEN ALERTED TO THIS MALFUNCTION THIS COULD POTENTIALLY RESULT IN FALSE NEGATIVE. FALSE NEGATIVES COULD POTENTIALLY CAUSE INJURY TO THE PT BY FAILING TO IDENTIFY A NERVE. AT THIS TIME WE ARE UNABLE TO CONFIRM WHETHER THE CUSTOMER RECEIVED ANY ERROR MESSAGES OR ALERTS OF THE FAULT. THIS REPORTED EVENT HAS NO REFERENCE TO AN ALARM OR WARNING, THEREFORE, IT MUST BE ASSUMED THAT AN ALARM OR WARNING WENT UNDETECTED OR DID NOT OCCUR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 108217 | AXON PATIENT MODULE | GWF- STIMULATOR, ELECTRICAL, EVOKED RESPONSE | GWF | MEDTRONIC XOMED, INC. | 945OPM660 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |