FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3010952 · Received March 14, 2013

Report

Report Number
1720753-2013-03779
Event Type
Malfunction
Date Received
March 14, 2013
Date of Event
February 22, 2013
Report Date
March 14, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE X-RAY CONTROLLER BOARD AND PERFORMED A FILAMENT CALIBRATION. THE SYS OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS DISPLAYED AN ERROR AND LOCKED UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
108322 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1