STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2024168-2013-01561
- Event Type
- Injury
- Date Received
- March 19, 2013
- Date of Event
- January 16, 2013
- Report Date
- March 1, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). ESTIMATED DATE OF OCCURRENCE, EXACT DATE WAS BETWEEN (B)(6) 2013. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE SE DEVICE AFTER A PERCUTANEOUS CORONARY INTERVENTIONAL PROCEDURE. REPORTEDLY, THE CLIP "SLID UP" THE DEVICE AND DEPLOYMENT WAS STOPPED. THE SAFETY RELEASE MECHANISM WAS USED TO REMOVE THE DEVICE. HEMOSTASIS WAS ACHIEVED BY APPLYING MANUAL ARTERIAL COMPRESSION. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE STARCLOSE SE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 114738 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | IMPLANTABLE CLIP | MGB | AV-TEMECULA-CT | 21002K1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |