ACCU-CHEK ® SPIRIT COMBO
Report
- Report Number
- 2183996-2013-00406
- Event Type
- Malfunction
- Date Received
- March 19, 2013
- Date of Event
- March 7, 2013
- Report Date
- May 29, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE COMPLAINT CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATION REGARDING THE CUSTOMER'S ALLEGATION. THE DELIVERY ACCURACY OF THE INSULIN PUMP WAS TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE TECHNICAL INVESTIGATION GAVE NO EVIDENCE THAT THE DEVICE DID CAUSE OR CONTRIBUTE TO THE CONDITION REPORTED BY THE PATIENT.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
PATIENT WAS AT A FACILITY FOR DIABETES READJUSTMENT, AND SHE REVIEWED HER WRITTEN BASAL AND BOLUS AMOUNTS AND FOUND THE AMOUNT WAS NOT CONSISTENT WITH THE INFUSION DEVICE HISTORY (UP TO 2, I.E. DIFFERENCE). SHE BELIEVES THE INSULIN DELIVERY OF THE INFUSION DEVICE IS INACCURATE. THE INFUSION DEVICE WAS REQUESTED FOR EVALUATION. NO PHYSIOLOGICAL EFFECTS WERE REPORTED, AND SHE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 114438 | ACCU-CHEK ® SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 047 YR |