FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT COMBO

MDR report key: 3010532 · Received March 19, 2013

Report

Report Number
2183996-2013-00404
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
March 1, 2013
Report Date
May 22, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE COMPLAINT CANNOT BE VERIFIED. THE PRODUCT MEETS THE SPECIFICATION REGARDING THE CUSTOMER'S ALLEGATION. THE DELIVERY ACCURACY OF THE INSULIN PUMP WAS TESTED WITHIN THE PUMP DRIVE TEST ON THE DIAGNOSTIC TEST SYSTEM AND MEETS THE SPECIFICATIONS. THE TECHNICAL INVESTIGATION GAVE NO EVIDENCE THAT THE DEVICE DID CAUSE OR CONTRIBUTE TO THE CONDITION REPORTED BY THE PATIENT.

Description of Event or Problem · 1

PATIENT EXPERIENCED HYPERGLYCEMIA OF 200-400 MG/DL FOR 1 WEEK. SHE CHANGED THE ACCESSORIES AND DELIVERED INSULIN VIA THE INFUSION DEVICE, BUT THIS DID NOT LOWER HER BLOOD GLUCOSE. HER INFUSION SITE WAS PAINFUL, AND SHE INITIALLY BELIEVED THIS WAS THE CAUSE. SHE SWITCHED TO A DIFFERENT TYPE OF INFUSION SET, BUT THIS DID NOT RESOLVE THE ISSUE. SHE NOW BELIEVES THE INSULIN DELIVERY OF THE INFUSION DEVICE IS TOO LOW. THE INFUSION DEVICE WAS REQUESTED FOR EVALUATION. SHE DID NOT REQUIRE TREATMENT FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113181 ACCU-CHEK ® SPIRIT COMBO INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1 020 YR