CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-01284
- Event Type
- Injury
- Date Received
- March 14, 2013
- Date of Event
- January 16, 2013
- Report Date
- February 21, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS INCLUDED IN A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULT: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT # 1627487-2013-1285. IT WAS REPORTED THE PT WAS EXPERIENCING SOME SLIGHT POCKET HEATING AFTER 30 MINUTES OF CHARGING. A REPLACEMENT CHARGER WAS SENT TO THE PT TO ADDRESS THE ISSUE AND HE WAS ADVISED TO CHARGE FOR SHORTER INTERVALS MORE FREQUENTLY TO ALLEVIATE THE POCKET HEATING. FOLLOW UP INFO IDENTIFIED THE ISSUE IS NOW RESOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 108221 | CHARGING SYSTEM | SCS CHARGER | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3701 | 3142666 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Required Intervention | IMPLANT DATE:| SCS LEAD: MODEL 3186 |