CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-13293
- Event Type
- Injury
- Date Received
- March 14, 2013
- Date of Event
- June 1, 2009
- Report Date
- February 21, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR. REPORT: 1627487-2013-13294. IT WAS REPORTED THE PT HAD EXPERIENCED POCKET HEATING WHILE CHARGING. THE PT STATED THIS HAS OCCURRED SINCE HE WAS FIRST IMPLANTED WITH HIS SCS SYSTEM. FOLLOW-UP INFORMATION IDENTIFIED THE PT WAS GIVEN A NEW LOW ENERGY CHARGER, HOWEVER, THE NEW CHARGER WAS UNABLE TO COMMUNICATE WITH THE PT'S IPG. THE PT'S OLD CHARGER DOES COMMUNICATE WITH THE PT'S IPG BUT SOMETIMES FLASHES A SLOW GREEN LIGHT WHEN CHARGING. A NEW CHARGING WAND HAS BEEN SENT TO THE PT TO TRY AND SEE IF IT RESOLVES THE ISSUE. FOLLOW-UP PENDING. ON (B)(4) 2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SET FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 108346 | CHARGING SYSTEM | SCS CHARGER | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | 2791323 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Required Intervention | IMPLANT DATE:| SCS LEAD: MODEL: 3219 |