FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 3010515 · Received March 14, 2013

Report

Report Number
1627487-2013-13293
Event Type
Injury
Date Received
March 14, 2013
Date of Event
June 1, 2009
Report Date
February 21, 2013
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. MANUFACTURER'S EVALUATION: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REFERENCE MFR. REPORT: 1627487-2013-13294. IT WAS REPORTED THE PT HAD EXPERIENCED POCKET HEATING WHILE CHARGING. THE PT STATED THIS HAS OCCURRED SINCE HE WAS FIRST IMPLANTED WITH HIS SCS SYSTEM. FOLLOW-UP INFORMATION IDENTIFIED THE PT WAS GIVEN A NEW LOW ENERGY CHARGER, HOWEVER, THE NEW CHARGER WAS UNABLE TO COMMUNICATE WITH THE PT'S IPG. THE PT'S OLD CHARGER DOES COMMUNICATE WITH THE PT'S IPG BUT SOMETIMES FLASHES A SLOW GREEN LIGHT WHEN CHARGING. A NEW CHARGING WAND HAS BEEN SENT TO THE PT TO TRY AND SEE IF IT RESOLVES THE ISSUE. FOLLOW-UP PENDING. ON (B)(4) 2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SET FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
108346 CHARGING SYSTEM SCS CHARGER LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 2791323

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention IMPLANT DATE:| SCS LEAD: MODEL: 3219