FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 3010352 · Received March 19, 2013

Report

Report Number
2024168-2013-01559
Event Type
Injury
Date Received
March 19, 2013
Date of Event
February 20, 2013
Report Date
February 27, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P050007
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). THE OTHER STARCLOSE SE DEVICE REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH MFR NUMBER. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EVENT WAS CONFIRMED. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. ANALYSIS ALSO REVEALED THAT THE STARCLOSE SE DEVICE MAY HAVE BEEN USED AFTER THE LABELED EXPIRATION DATE PER THE REPORTED DATE OF OCCURRENCE. PER THE INSTRUCTIONS FOR USE - GRAPHICAL SYMBOLS FOR MEDICAL DEVICE LABELING - USE BY AS LABELED. SEVEN UNUSED STERILE REPRESENTATIVE SAMPLES WITH THE SAME PART AND LOT NUMBERS WERE RETURNED FOR EVALUATION. FUNCTIONAL TESTING WAS PERFORMED ON THE SEVEN RETURNED DEVICES AND THE DEVICES PASSED WITH ACCEPTABLE RESULTS. NO MANUFACTURING ISSUES OR ABNORMAL OBSERVATIONS WERE DETECTED. A CAUSE FOR THE REPORTED EXPERIENCE COULD NOT BE DETERMINED BASED ON THE INVESTIGATION FINDINGS FROM THE TESTED SAMPLES.

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MEDWATCH REPORT, THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED: REPORTEDLY, THE PHYSICIAN ATTEMPTED TO INTRODUCE THE STARCLOSE SE EXCHANGE SHEATH, BUT IT BENT. A SECOND STARCLOSE SE WAS USED WITH THE SAME RESULTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE SE DEVICE AFTER A SUPERIOR FEMORAL ARTERY DILATATION INTERVENTIONAL PROCEDURE. REPORTEDLY, THE STARCLOSE SE INTRODUCER BENT AND IT WAS NOT POSSIBLE TO ADVANCE IN THE ARTERY. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN REPORTEDLY IS IN-TRAINING IN THE USE OF THE STARCLOSE SE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113547 STARCLOSE SE VASCULAR CLOSURE SYSTEM IMPLANTABLE CLIP MGB AV-TEMECULA-CT 020266H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6FHEPARIN