FDA Adverse Event Malfunction Summary report: N

FLEXCATH STEERABLE SHEATH

MDR report key: 3010283 · Received March 19, 2013

Report

Report Number
3002648230-2013-00045
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
February 27, 2013
Report Date
February 27, 2013
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. RESULTS OF EVALUATION OF RETURNED DEVICE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE VISUAL INSPECTION SHOWED THAT THE SHAFT WAS INTACT WITH NO APPARENT ISSUES. THE REPORTED LEAK WAS REPRODUCED WHEN A CATHETER WAS INTRODUCED THROUGH THE SHEATH. DISSECTION SHOWED THE HEMOSTATIC VALVE WAS LEAKING; VALVE WAS TORN. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

DURING THE PROCEDURE, FLUID WAS LEAKING OUT OF THE SHEATH'S VALVE. THE PHYSICIAN DECIDED NOT TO EXCHANGE THE SHEATH AND THE PROCEDURE WAS COMPLETED WITH THIS SHEATH. THERE WERE NO PATIENT SYMPTOMS/INJURIES RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114773 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12

Patients

Seq Age Sex Outcome Treatment
1