FLEXCATH STEERABLE SHEATH
Report
- Report Number
- 3002648230-2013-00045
- Event Type
- Malfunction
- Date Received
- March 19, 2013
- Date of Event
- February 27, 2013
- Report Date
- February 27, 2013
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION. RESULTS OF EVALUATION OF RETURNED DEVICE WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.
THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE VISUAL INSPECTION SHOWED THAT THE SHAFT WAS INTACT WITH NO APPARENT ISSUES. THE REPORTED LEAK WAS REPRODUCED WHEN A CATHETER WAS INTRODUCED THROUGH THE SHEATH. DISSECTION SHOWED THE HEMOSTATIC VALVE WAS LEAKING; VALVE WAS TORN. A FIELD ACTION WAS INITIATED IN JULY 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
DURING THE PROCEDURE, FLUID WAS LEAKING OUT OF THE SHEATH'S VALVE. THE PHYSICIAN DECIDED NOT TO EXCHANGE THE SHEATH AND THE PROCEDURE WAS COMPLETED WITH THIS SHEATH. THERE WERE NO PATIENT SYMPTOMS/INJURIES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 114773 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |