FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3010181 · Received March 19, 2013

Report

Report Number
2210968-2013-02575
Event Type
Malfunction
Date Received
March 19, 2013
Date of Event
February 28, 2013
Report Date
February 28, 2013
Manufacturer
ETHICON, INC,
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE FAILED TO ROTATE DURING THE EVALUATION; IT IS LIKELY THAT THE ACCUMULATION OF BLOOD, BODY FLUIDS, AND TISSUE PREVENTED THE DEVICE FROM CONTINUING TO OPERATE AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE DEVICE ACTIVATED PROPERLY AT THE BEGINNING, THEN, THE BLADE STOPPED ROTATING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114892 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC, NA MT216993

Patients

Seq Age Sex Outcome Treatment
1