OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-05562
- Event Type
- Malfunction
- Date Received
- March 19, 2013
- Report Date
- February 19, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (06/27/2013)-DEVICE EVALUATION: THE LAY USER/PATIENT¿S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 06/20/2013 WITH THE FOLLOWING FINDINGS: THE METER INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE METER¿S CAPACITOR WAS FOUND TO BE DEFECTIVE. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE METER ASSOCIATED WITH THIS COMPLAINT HAS BEEN RETURNED TO LIFESCAN; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.
ON (B)(4) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING BATTERY INDICATOR ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 113634 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3304708 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |