FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 3010019 · Received March 18, 2013

Report

Report Number
1644487-2013-00726
Event Type
Malfunction
Date Received
March 18, 2013
Date of Event
January 24, 2013
Report Date
February 19, 2013
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PRODUCT ANALYSIS OF THE COMPUTER AND FLASHCARD WAS COMPLETED. NO ANOMALIES ASSOCIATED WITH THE HANDHELD COMPUTER WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE HANDHELD PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO SPECIFICATIONS. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.

Description of Event or Problem · 1

REPORTER INDICATED A DELL X5 COMPUTER WITH 8.0 SOFTWARE WAS FREEZING ON THE "INTERROGATION SUCCESSFUL" SCREEN FOR THE PREVIOUS 3 WEEKS. IT WAS NOT CLEAR HOW LONG THE SCREEN WAS FREEZING. THE COMPUTER AND FLASHCARD HAVE BEEN RETURNED AND ARE PENDING PRODUCT ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
111937 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 839583

Patients

Seq Age Sex Outcome Treatment
1