FDA Adverse Event Other Summary report: N

KARL STORZ

MDR report key: 3009244 · Received March 12, 2013

Report

Report Number
9610617-2013-00009
Event Type
Other
Date Received
March 12, 2013
Date of Event
January 30, 2013
Report Date
March 11, 2013
Manufacturer
KARL STORZ GMBH & CO. KG
Product Code
FCL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE JAWS OF THE BIOPSY FORCEPS ARE STUCK IN OPEN POSITION; POSSIBLE DAMAGE TO THE DRAW BAR DUE TO STRESS OVERLOAD. WE SHOW NO OTHER COMPLAINTS FOR THIS ISSUE WITH THIS PRODUCT.

Description of Event or Problem · 1

ALLEGEDLY, THE DOCTOR WAS PERFORMING A VENTRICULAR BIOPSY AND WHEN SHE SENT TO GRASP TISSUE, THE JAWS WOULD NOT CLOSE. THE DOCTOR EXPERIENCED DIFFICULTIES IN REMOVING THE INSTRUMENT THAT ADDED AN HOUR TO THE PROCEDURE. INSTRUMENT WAS REPLACED AND PROCEDURE COMPLETED. PATIENT CONDITION IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104782 KARL STORZ BIOPSY FORCEPS FCL KARL STORZ GMBH & CO. KG 11161KA TY

Patients

Seq Age Sex Outcome Treatment
1 Other