FDA Adverse Event Malfunction Summary report: N

INTRAMEDULLARY ARTHRODESIS IMPLANT SMART TOE II 19MM / 10° ANGLE FOR PIP ARTHRO

MDR report key: 3007524 · Received March 15, 2013

Report

Report Number
0008031020-2013-00055
Event Type
Malfunction
Date Received
March 15, 2013
Date of Event
February 20, 2013
Report Date
February 20, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-SELZACH
Product Code
HTY
PMA / PMN Number
K112197
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE REPORTED EVENT THAT THE IMPLANT DID NOT EXPAND COULD NOT BE CONFIRMED SINCE THE RETURNED DEVICE IS CONFORMING TO SPECIFICATION. THEREFORE, THE CASE COULD BE CLASSIFIED AS A NON ISSUE. WITH THE LABELING REVIEW, IT WAS VERIFIED THAT THE OP TECH READS: IMPLANTS ARE DELIVERED STERILE - THEY NEED TO BE PLACED IN THE FREEZER (0 DEGREES C /32 DEGREES F, OR BELOW) FOR 2 HOURS OR MORE PRIOR TO IMPLANTATION. NOTE: TO FACILITATE THE SHAPE MEMORY PROCESS, THE PHALANGES CAN BE BATHED IN A WARM STERILE SOLUTION, BETWEEN 37 DEGREES C (98.6 DEGREES F) TO 40 DEGREES C (104 DEGREES F). WE CAN CONCLUDE THE ROOT CAUSE OF THIS COMPLAINT IS RELATED TO USE. THERE IS NO SIMILAR COMPLAINT REPORTED WITHIN THE SAME LOT NUMBER. ONE DEVIATION FROM THE SPECIFICATION WAS NOTED BUT NOT RELATED TO THE REPORTED FAILURE (FOREIGN BODIES ON 2 PARTS AFTER STERILIZATION, 2 NON CONFORM DEVICES WERE SCRAPPED (B)(4)).

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

SURGEON SAID THE SMART TOE DID NOT FUNCTION PROPERLY. THEY OPENED ANOTHER ONE AND THAT ONE WORKED FINE.

Description of Event or Problem · 1

SURGEON SAID THE SMART TOE DID NOT FUNCTION PROPERLY. THEY OPENED ANOTHER ONE AND THAT ONE WORKED FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109822 INTRAMEDULLARY ARTHRODESIS IMPLANT SMART TOE II 19MM / 10° ANGLE FOR PIP ARTHRO IMPLANT HTY STRYKER OSTEOSYNTHESIS-SELZACH F003279PBBM

Patients

Seq Age Sex Outcome Treatment
1 Other