FDA Adverse Event Injury Summary report: N

M2A-MAGNUM MOD HD SZ 44MM

MDR report key: 3007495 · Received March 15, 2013

Report

Report Number
0001825034-2013-00630
Event Type
Injury
Date Received
March 15, 2013
Date of Event
February 16, 2013
Report Date
February 20, 2013
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DIMENSIONAL EVALUATION FOUND COMPONENT TO BE WITHIN APPROPRIATE DESIGN SPECIFICATION. EXAMINATION OF THE RETURNED DEVICES FOUND LITTLE EVIDENCE OF BONY INGROWTH ON THE CUP. THE HEAD AND CUP APPEARED TO SHOW EVIDENCE OF SUBLUXATION OF THE JOINT AND SCRATCHING OF THE BEARING SURFACES. BASED UPON THE APPEARANCE OF THE PARTS, WEAR RATES DID NOT APPEAR TO BE EXCESSIVE. THIS REPORT IS NUMBER 2 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2013-00629-1 / 00631-1).

Additional Manufacturer Narrative · 1

EVALUATION IN PROCESS BUT NOT YET COMPLETE. UPON COMPLETION OF EVALUATION, A FOLLOW UP REPORT WILL BE SENT TO THE FDA. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS, NUMBER 1 STATES, "MATERIAL SENSITIVITY REACTIONS." NUMBER 14 STATES, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." AND NUMBER 4 STATES, "LOOSENING OR MIGRATION OF THE IMPLANTS MAY OCCUR DUE TO LOSS OF FIXATION, TRAUMA, MALALIGNMENT, BONE RESORPTION, OR EXCESSIVE ACTIVITY." THIS REPORT IS NUMBER 2 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2013-00629 / 00631).

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT UNDERWENT A TOTAL HIP ARTHROPLASTY ON (B)(6) 2009. SUBSEQUENTLY, PATIENT WAS REVISED ON (B)(6) 2013 DUE TO PAIN, LOOSENING, AND PSEUDOTUMOR. THE CUP, HEAD AND TAPER ADAPTER WERE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
109211 M2A-MAGNUM MOD HD SZ 44MM PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 323220

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| R