ALENTI
Report
- Report Number
- 9611530-2013-00018
- Event Type
- Injury
- Date Received
- March 12, 2013
- Date of Event
- January 20, 2013
- Report Date
- February 13, 2013
- Manufacturer
- ARJO HOSPITAL EQUIPMENT AB
- Product Code
- FSA
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING FILED UNDER (B)(4) BY ARJOHUNTLEIGH (B)(4) ON BEHALF OF THE IMPORTER (B)(4). THIS IS THE FIRST REPORTABLE EVENT FOR A SIMILAR FAILURE MODE AND THE DEVICE (ARM SUPPORT DROP CAUSING FRACTURES). THERE IS NO TREND OBSERVED FOR REPORTABLE COMPLAINTS WITH THIS FAILURE MODE ON ALENTI DEVICE. WE WERE NOT ABLE TO FIND ANY DEFICIENCY WITH THE DEVICE. THIS MEANS THAT THE SHOWERING SYSTEM WAS UP TO SPECIFICATION WHEN THE EVENT TOOK PLACE. THE DEVICE WAS BEING USED FOR PATIENT HANDLING AND IN THAT WAY CONTRIBUTED TO THE EVENT. FROM OUR EVALUATION IT APPEARS A NUMBER OF USE ERRORS HAVE CAUSE THE EVENT, THE MOST RELEVANT USE ERROR BEING A FAILURE TO EVALUATE THE PERSON TO BE TRANSFERRED BEFORE USE: NEED FOR THE PROFESSIONAL CLINICAL ASSESSMENT PRIOR TO USE OF ALENTI DEVICE IS STATED IN THE INSTRUCTION FOR USE (04.CD.02/7 US/CA): 'WE RECOMMENDED THAT THE FACILITIES ESTABLISH REGULAR ASSESSMENT ROUTINES. CAREGIVERS SHOULD ASSESS EACH RESIDENT ACCORDING TO CRITERIA PRIOR TO USE [...], I.E. THE RESIDENT SHOULD UNDERSTAND AND RESPOND TO INSTRUCTIONS TO STAY SEATED IN AN UPRIGHT POSITION.' ADDITIONALLY IT IS WARNED: 'WARNING! WHEN POSITIONING THE LIFT HYGIENE CHAIR, DO NOT STAND IN THE DIRECTION OF THE MOVEMENT OF THE HAND REST. AN UNEXPECTED MOVEMENT OF THE ARMS UPWARDS, BY THE RESIDENT, CAN INJURY THE NURSE.' WE HAVE NOT BEEN ABLE TO FIND ANY CONTRIBUTING MANUFACTURING ANOMALIES. THE ROOT CAUSE OF THE COMPLAINT WAS FOUND TO BE A USE ERROR AS THE RECEIVED INFORMATION AND OUR EVALUATION AS DESCRIBED ABOVE ARE SHOWING THAT IF THE IFU SAFETY WARNINGS ARE FOLLOWED THERE WILL BE NO PATIENT OR CAREGIVER RISK.
ACCORDING TO THE DETAILS REPORTED BY THE ARJOHUNTLIEGH REPRESENTATIVE, THE FOLLOWING OCCURRED: 'CAREGIVER ATTEMPTING TO GIVE BATH. RESIDENT WAS FINE GETTING INTO BATH, AS SOON AS WATER TURNED ON, HE FREAKED OUT SCREAMING AND YELLING AND FLUNG SAFETY BAR OF TUB CHAIR OFF WHICH LANDED ON SIDE OF LEFT WRIST OF CAREGIVER.'
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104077 | ALENTI | HYGIENE AND POOL LIFTERS | FSA | ARJO HOSPITAL EQUIPMENT AB | CDB8153-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |