FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3004028 · Received March 14, 2013

Report

Report Number
1416980-2013-06109
Event Type
Malfunction
Date Received
March 14, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CAUSE OF THE EVENT WAS USE ERROR. THERE WAS NO REPORTED DEVICE MALFUNCTION; THEREFORE, NO FURTHER INVESTIGATION WILL BE PERFORMED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THE LABELING INSTRUCTS THE PATIENT TO ENSURE ALL CLAMPS ARE CLOSED ON UNUSED LINES.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240, (AIR IN TUBING) WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE. THE HOME PATIENT (HP) STATED THAT THERE WERE NO LEAKS OR WET LINES. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE PATIENT LINE HAD BEEN PROPERLY PRIMED AND NO PATIENT EXTENSIONS WERE IN USE. THE PATIENT HAD NOT DISCONNECTED PRIOR TO THE ALARM. THE BAGS WERE PROPERLY CONNECTED AND THERE WAS AN OPEN CLAMP ON AN UNUSED LINE. THERE WAS NOTHING UNUSUAL NOTED ABOUT THE SUPPLIES, AND THEY HAD NOT BEEN DAMAGED BY AN OUTLET PORT CLAMP OR ASSIST DEVICE. THE TSR EXPLAINED THE ALARM AND CLEARED IT. THE TSR ADVISED THE HP TO CALL THE NURSE FOR FURTHER MEDICAL INSTRUCTIONS. THERE WAS A PATIENT INVOLVED, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107783 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 62 YR HOMECHOICE