FDA Adverse Event
Malfunction
Summary report: N
TOTAL CARE BED
MDR report key: 3004017
·
Received February 19, 2013
Report
- Report Number
- 1824206-2013-01015
- Event Type
- Malfunction
- Date Received
- February 19, 2013
- Date of Event
- September 6, 2012
- Report Date
- September 6, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND SIGNS OF FLUID INGRESS. THE TECHNICIAN USED AN ANTIOXIDANT SPRAY ON THE SIDE RAIL ELECTRICAL CONTACTS TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE ACCOUNT ALLEGED THAT THE SIDE RAILS HAVE NO FUNCTIONS. NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 71612 | TOTAL CARE BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |