FDA Adverse Event Injury Summary report: N

UNKNOWN IMPLANTABLE NEUROSTIMULATOR

MDR report key: 3003879 · Received March 14, 2013

Report

Report Number
3007566237-2013-00776
Event Type
Injury
Date Received
March 14, 2013
Report Date
February 21, 2013
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
EZW
PMA / PMN Number
P970004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF LEAD (1 OF 2) FOUND THE STIMULATION LEAD BODY CONDUCTOR BROKEN AT OR NEAR TINES. ALL CONDUCTORS WERE BROKEN. ANALYSIS OF LEAD (2 OF 2) FOUND THE STIMULATION LEAD BODY CONDUCTOR BROKEN AT OR NEAR TINES. ALL CONDUCTORS WERE BROKEN.

Additional Manufacturer Narrative · 1

PRODUCT ID: 3889, SERIAL# UNKNOWN, IMPLANTED: (B)(6) 2011, EXPLANTED: (B)(6) 2013, PRODUCT TYPE: LEAD. PRODUCT ID: 3889, SERIAL# UNKNOWN, IMPLANTED: (B)(6) 2011, EXPLANTED: (B)(6) 2013, PRODUCT TYPE: LEAD. (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT EXPERIENCED LESS THAN 50 PERCENT THERAPY RELIEF AND HIGH IMPEDANCES WERE REPORTED. IT WAS ALSO REPORTED THE PATIENT HAD THE LEADS IMPLANTED NEAR THE PUDENDUS NERVE DUE TO PELVIC PAIN. THE PATIENT WAS REPORTED TO HAVE BIKED SEVERAL TIMES BEFORE A LOSS OF THERAPY OCCURRED. IT WAS ALSO REPORTED THE PHYSICIAN TRIED TO REPROGRAM THE PATIENT'S DEVICE WITHOUT SUCCESS AND IMPEDANCE MEASUREMENTS SHOWED HIGH IMPEDANCES, GREATER THAN 4,000 OHMS. IT WAS REPORTED THE PATIENT'S LEADS WERE EXPLANTED AND DURING THE EXPLANT PROCEDURE THE LEADS WERE CUT AND THE PROXIMAL PART OF THE LEADS WERE LEFT IMPLANTED IN THE EXTENSION CONNECTOR. IT WAS ALSO REPORTED THE EXTENSION AND BATTERY WERE LEFT IMPLANTED AND THE PATIENT RECOVERED WITH NO INJURY OR ADVERSE EVENT. ADDITIONAL INFORMATION HAS BEEN REQUESTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107607 UNKNOWN IMPLANTABLE NEUROSTIMULATOR STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE EZW MEDTRONIC NEUROMODULATION NEU_INS_STIMULATOR

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention