FDA Adverse Event
Malfunction
Summary report: N
VERSACARE BED
MDR report key: 3003597
·
Received February 11, 2013
Report
- Report Number
- 1824206-2013-00953
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- February 6, 2013
- Report Date
- February 6, 2013
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN REPLACED THE AC POWER CORD PLUG TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
INFO RECEIVED INDICATED THE BED'S GROUND IMPEDANCE (RESISTANCE) IS OUT OF SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 59212 | VERSACARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |