FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 3003517
·
Received February 11, 2013
Report
- Report Number
- 1811605-2013-00012
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- January 28, 2013
- Report Date
- January 28, 2013
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
IT WAS REPORTED THE FABRIC FOUNDATION OF THE UNIT WAS BURNED BY AN INTERNAL COMPONENT. OUR EXAMINATION REVEALED THAT ONE OF THE TERMINALS HAD BEEN BROKEN NEAR A THERMOSTAT, RESULTING IN A SPARK WHICH HAD CAUSED A 1/4" BURN HOLE IN ONE SIDE (BOTH SIDES) OF THE FABRIC FOUNDATION. THIS TYPE OF INCIDENT IS NORMALLY CAUSED BY MISUSE ON THE PART OF THE CONSUMER, BUT THERE WERE NO ADDITIONAL SIGNS OF BROKEN COMPONENTS OR MISUSE. WE WERE UNABLE TO DETERMINE THE CAUSE OF THIS REPORT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE UNIT HAD A SMALL BURN HOLE AND WOULD NOT WORK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 60385 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 156 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |