FDA Adverse Event Malfunction Summary report: N

THERMOPHORE AUTOMATIC MOIST HEAT PACK

MDR report key: 3003517 · Received February 11, 2013

Report

Report Number
1811605-2013-00012
Event Type
Malfunction
Date Received
February 11, 2013
Date of Event
January 28, 2013
Report Date
January 28, 2013
Manufacturer
BATTLE CREEK EQUIPMENT COMPANY
Product Code
IRT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THE FABRIC FOUNDATION OF THE UNIT WAS BURNED BY AN INTERNAL COMPONENT. OUR EXAMINATION REVEALED THAT ONE OF THE TERMINALS HAD BEEN BROKEN NEAR A THERMOSTAT, RESULTING IN A SPARK WHICH HAD CAUSED A 1/4" BURN HOLE IN ONE SIDE (BOTH SIDES) OF THE FABRIC FOUNDATION. THIS TYPE OF INCIDENT IS NORMALLY CAUSED BY MISUSE ON THE PART OF THE CONSUMER, BUT THERE WERE NO ADDITIONAL SIGNS OF BROKEN COMPONENTS OR MISUSE. WE WERE UNABLE TO DETERMINE THE CAUSE OF THIS REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE UNIT HAD A SMALL BURN HOLE AND WOULD NOT WORK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
60385 THERMOPHORE AUTOMATIC MOIST HEAT PACK ELECTRIC MOIST HEAT PACK IRT BATTLE CREEK EQUIPMENT COMPANY 156