12.0MM/8.0MM PROTECTION SLEEVE 188MM
Report
- Report Number
- 8030965-2013-10440
- Event Type
- Malfunction
- Date Received
- February 28, 2013
- Date of Event
- January 29, 2011
- Report Date
- January 29, 2011
- Manufacturer
- SYNTHES GMBH
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE (S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILLING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. SUBJECT DEVICE HAS BEEN REC'D AND EVALUATED. THE MFG RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. ALL REC'D DEVICES SHOWED MARKS OF FORCIBLE USE. THE PROTECTION SLEEVES HAD VISIBLE MARKS OF MISUSE CAUSED BY FORCIBLE ATTEMPTS TO STRIKE THEM INTO THE DEFORMED BORES OF THE AIMING ARMS. THE INTACT AND MEASURABLE DIAMETERS DID FULLY MEET TO OUR SPECIFICATIONS. THIS COMPLAINT IS DEEMED INVALID FROM A MFG STANDPOINT.
DURING A LATERAL ENTRY FEMORAL NAIL PROCEDURE FOR PROXIMAL FEMUR FRACTURE, THE PROTECTION SLEEVE DID NOT FIT THROUGH THE 120 DEGREE ANTE-GRADE HOLE (FOR THE LATERAL ENTRY FEMORAL NAIL) IN THE STANDARD INSERTION HANDLE. THE SURGEON TRIED THREE DIFFERENT SLEEVES AND THREE DIFFERENT HANDLES FROM DIFFERENT SETS. HE HAD TO AGGRESSIVELY MALLET THE SLEEVE INTO THE HANDLES AND USE STERILE MINERAL OIL. ONE OF THE SLEEVES WENT HALF WAY IN, AND WAS MALLETED AGGRESSIVELY. THE SURGEON COMPLETED THE PROCEDURE SUCCESSFULLY. THE SURGEON COMPLETED THE PROCEDURE SUCCESSFULLY. THIS EVENT DID NOT IMPOSE ANY HARM ON THE PT. THIS INCIDENT EXTENDED THE SURGERY BY APPROXIMATELY 30 MINUTES. THIS COMPLAINT IS ON THE FIRST PROTECTION SLEEVE. THIS REPORT IS 4 OF 6 FOR THE SAME EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87168 | 12.0MM/8.0MM PROTECTION SLEEVE 188MM | LXH | SYNTHES GMBH | 4786948 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |