FDA Adverse Event Malfunction Summary report: N

12.0MM/8.0MM PROTECTION SLEEVE 188MM

MDR report key: 3003480 · Received February 28, 2013

Report

Report Number
8030965-2013-10440
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
January 29, 2011
Report Date
January 29, 2011
Manufacturer
SYNTHES GMBH
Product Code
LXH
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE (S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILLING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. SUBJECT DEVICE HAS BEEN REC'D AND EVALUATED. THE MFG RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. ALL REC'D DEVICES SHOWED MARKS OF FORCIBLE USE. THE PROTECTION SLEEVES HAD VISIBLE MARKS OF MISUSE CAUSED BY FORCIBLE ATTEMPTS TO STRIKE THEM INTO THE DEFORMED BORES OF THE AIMING ARMS. THE INTACT AND MEASURABLE DIAMETERS DID FULLY MEET TO OUR SPECIFICATIONS. THIS COMPLAINT IS DEEMED INVALID FROM A MFG STANDPOINT.

Description of Event or Problem · 1

DURING A LATERAL ENTRY FEMORAL NAIL PROCEDURE FOR PROXIMAL FEMUR FRACTURE, THE PROTECTION SLEEVE DID NOT FIT THROUGH THE 120 DEGREE ANTE-GRADE HOLE (FOR THE LATERAL ENTRY FEMORAL NAIL) IN THE STANDARD INSERTION HANDLE. THE SURGEON TRIED THREE DIFFERENT SLEEVES AND THREE DIFFERENT HANDLES FROM DIFFERENT SETS. HE HAD TO AGGRESSIVELY MALLET THE SLEEVE INTO THE HANDLES AND USE STERILE MINERAL OIL. ONE OF THE SLEEVES WENT HALF WAY IN, AND WAS MALLETED AGGRESSIVELY. THE SURGEON COMPLETED THE PROCEDURE SUCCESSFULLY. THE SURGEON COMPLETED THE PROCEDURE SUCCESSFULLY. THIS EVENT DID NOT IMPOSE ANY HARM ON THE PT. THIS INCIDENT EXTENDED THE SURGERY BY APPROXIMATELY 30 MINUTES. THIS COMPLAINT IS ON THE FIRST PROTECTION SLEEVE. THIS REPORT IS 4 OF 6 FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87168 12.0MM/8.0MM PROTECTION SLEEVE 188MM LXH SYNTHES GMBH 4786948

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention