FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3003428 · Received February 13, 2013

Report

Report Number
1314492-2013-00099
Event Type
Malfunction
Date Received
February 13, 2013
Date of Event
January 15, 2013
Report Date
January 16, 2013
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BAXTER EVALUATED THE DEVICE IN QUESTION FOR 24 HRS AND FOUND ALL KEYS TO PERFORM AS EXPECTED AND NO KEYS RESULTED IN AN INCORRECT RESPONSE OR DOUBLE ENTRY. THE KEYPAD WAS DELAMINATED AND EXAMINED UNDER A MICROSCOPE AND NO FAILURES WERE EVIDENT ON THE KEYPAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHEN THE OK KEY ON A PUMP KEYPAD IS SELECTED, THE DEVICE ¿SHUTS OFF;¿ AND THE 3RD SOFT KEY AND THE #2 KEY ARE INOPERABLE. IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.