FDA Adverse Event Malfunction Summary report: N

OT ULTRALINK METER

MDR report key: 3003235 · Received March 14, 2013

Report

Report Number
3008382007-2013-04954
Event Type
Malfunction
Date Received
March 14, 2013
Report Date
February 15, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K073231
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE LIFESCAN PRODUCT HAS BEEN RETURNED TO LIFESCAN. EVALUATION HAS NOT YET BEEN COMPLETED. AFTER THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILLED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4): THE METER WAS FOUND TO HAVE SPC PINS 1-3 LIFTED HIGH. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING APPLY SAMPLE - METER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
107951 OT ULTRALINK METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3278630

Patients

Seq Age Sex Outcome Treatment
1 16 YR