FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 SYNCHRON® SYSTEM

MDR report key: 3003019 · Received March 13, 2013

Report

Report Number
2050012-2013-00170
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) NOTED BUBBLES IN THE RATIO PUMP, DEBRIS ON THE LOWER SEALS, AND VERTICAL GROOVES ON THE PISTON DUE TO WEAR. THE FSE REPLACED THE RATIO PUMP AND RESOLVED THE ISSUE. THE FSE ADDITIONALLY REPLACED THE CALCIUM ELECTRODE TIP TO RESOLVE CALIBRATION ISSUES; NO CALCIUM RESULTS WERE IMPACTED. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ELEVATED SODIUM RESULTS FOR QUALITY CONTROL (QC) AND TWO PATIENTS' SAMPLES INVOLVING THE UNICEL DXC 600 SYNCHRON SYSTEM. THE CUSTOMER STATED SODIUM QC AND PATIENT RESULTS WERE FIVE TO SIX POINTS HIGHER COMPARED TO AN ALTERNATE INSTRUMENT. CONTROL LEVEL 1 RECOVERED AT 159 MMOL/L. NO ERRONEOUS PATIENT RESULTS WERE RELEASED OUT OF THE LABORATORY. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104960 UNICEL® DXC 600 SYNCHRON® SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1