UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM
Report
- Report Number
- 1061932-2013-00397
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- February 14, 2013
- Report Date
- February 14, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K081930
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT. THE FSE HAD OBSERVED THAT THE INSTRUMENT WAS SPLATTERING BLOOD ONTO THE CASSETTES AND THE SPECIMEN TRANSPORT MODULE (STM). THE FSE ADJUSTED THE PROBE WASH COLLAR UNTIL IT WAS FLUSH WITH THE OPENING IN THE PROBE AND CLEANED THE CASSETTES AND THE STM. THE FSE RAN SHUTDOWN AND DAILY CHECKS. THE INSTRUMENT RAN WITHOUT ANY ISSUES. PER LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM SPLATTERED BLOOD IN MULTIPLE SPOTS THAT WERE GETTING ON THE CASSETTE LABELS. THERE WAS NO SPILL JUST SPLATTER. THE INSTRUMENT DID NOT GENERATE ANY ERROR MESSAGES OR FLAGS AS A RESULT OF THIS EVENT. THE CUSTOMER WAS WEARING PROPER PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT AND GLOVES AT THE TIME OF OCCURRENCE. THERE WAS NO CONTACT TO OPEN OR BROKEN SKIN OR MUCOUS MEMBRANES. NO ONE SOUGHT MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 105595 | UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | DXH 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |