FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3002996 · Received March 13, 2013

Report

Report Number
1061932-2013-00399
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPLACED THE TUBING THAT WAS LEAKING FLUID AROUND THE DIFFERENTIAL MIXING CHAMBER AT VL46, WHICH FIXED THE LEAK. THE UNIT IS OPERATIONAL. SERVICE WAS NOT DISPATCHED AS THE CUSTOMER CORRECTED THE ISSUE. PER LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT THERE WAS A LEAK AROUND THE DIFFERENTIAL MIXING CHAMBER AT VL46 OF THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VL46 CONTROLS THE RINSING OF THE LOWER CHAMBER OF THE FLOW CELL WHICH IS USED FOR BOTH DIFFERENTIAL AND RETICS PROCESSES. THE VOLUME OF FLUID WAS 7-8 MLS OF CLEAR FLUID ON THE COUNTER TOP. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT, GLOVES, AND EYE GLASSES DURING THIS EVENT. THERE WAS NO CONTACT TO OPEN OR BROKEN SKIN OR MUCOUS MEMBRANES. NO ONE SOUGHT MEDICAL ATTENTION. THERE WAS NO DEATH, INJURY OR AFFECT TO PATIENT TREATMENT. NO ERRONEOUS RESULTS WERE GENERATED.