PROGRASP FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-00819
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- November 13, 2012
- Report Date
- February 14, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT HAD A BROKEN WIRE AT THE TIP. THE ONE GRIP OPEN CABLE WAS FRAYED NEAR THE FORCE AMP PULLEY AND THE DISTAL IDLER PULLEY. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE PROGRASP FORCEPS INSTRUMENT HAD A BROKEN WIRE AT THE TIP OF THE INSTRUMENT. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104869 | PROGRASP FORCEPS INSTRUMENT | ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420093-08 | M11120703 665 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES |