PROGRASP FORCEPS INSTRUMENT

MDR report key: 3002970 · Received March 13, 2013

Report

Report Number
2955842-2013-00819
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
November 13, 2012
Report Date
February 14, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT HAD A BROKEN WIRE AT THE TIP. THE ONE GRIP OPEN CABLE WAS FRAYED NEAR THE FORCE AMP PULLEY AND THE DISTAL IDLER PULLEY. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE PROGRASP FORCEPS INSTRUMENT HAD A BROKEN WIRE AT THE TIP OF THE INSTRUMENT. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104869 PROGRASP FORCEPS INSTRUMENT ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M11120703 665

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES