TROCHANTERIC NAIL KIT, TI GAMMA3 Ø10X170MM X 125°
Report
- Report Number
- 0009610622-2013-00113
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- February 16, 2013
- Report Date
- February 16, 2013
- Manufacturer
- STRYKER OSTEOSYNTHESIS-KIEL
- Product Code
- HSB
- PMA / PMN Number
- K034223
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION SUMMARY: A REVIEW OF THE DHR REVEALED NO DEVIATION. THE REVIEW OF THE MANUFACTURING DOCUMENTS REVEALED NO DEVIATION DURING THE PACKAGING PROCESS. THE HAIR WAS FOUND AT THE RIM OF THE BLISTER POUCH ON THE GLUE AREA. THE TYVEK LID HAD BEEN PEELED OFF ALREADY BY THE USER. THE TYVEK LID SHOWS A CORRESPONDING IMPRINT THAT WAS CAUSED BY THE CAPTURED HAIR. THUS IT COULD BE CONCLUDED THAT THE HAIR GOT CAPTURED DURING THE SEALING PROCESS AT THE SUPPLIER.
IT WAS REPORTED THAT WHILE SCRUB NURSE WAS OPENING THE PACKAGE, A LONG RED HAIR WAS STUCK TO THE GLUE INSIDE AND ALSO STUCK OUT OF THE INNER BLISTER.
IT WAS REPORTED THAT WHILE SCRUB NURSE WAS OPENING THE PACKAGE, A LONG RED HAIR WAS STUCK TO THE GLUE INSIDE AND ALSO STUCK OUT OF THE INNER BLISTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 105716 | TROCHANTERIC NAIL KIT, TI GAMMA3 Ø10X170MM X 125° | IMPLANT | HSB | STRYKER OSTEOSYNTHESIS-KIEL | K700548 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other |