FDA Adverse Event Malfunction Summary report: N

MULTI-LINK 8

MDR report key: 3002874 · Received March 13, 2013

Report

Report Number
2024168-2013-01455
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 18, 2013
Report Date
February 18, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MAF
PMA / PMN Number
P020047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT SEPARATION WAS CONFIRMED. BASED ON ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES FOR THIS LOT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED STATED THAT THERE WAS NO RESISTANCE REMOVING THE DEVICE FROM THE PACKAGING HOOP. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE 2.75X15 MM MULTI-LINK 8 SHAFT SEPARATED DURING PREPARATION OF THE DEVICE AND COULD NOT BE USED. ANOTHER UNSPECIFIED DEVICE WAS USED SUCCESSFULLY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105912 MULTI-LINK 8 CORONARY STENT SYSTEM MAF AV-TEMECULA-CT 2012441

Patients

Seq Age Sex Outcome Treatment
1