FDA Adverse Event Injury Summary report: N

AVIATOR ASSY FOUR LEVEL PLATE SIZE 56

MDR report key: 3002828 · Received March 13, 2013

Report

Report Number
0009617544-2013-00072
Event Type
Injury
Date Received
March 13, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
STRYKER SPINE-FRANCE
Product Code
KWQ
PMA / PMN Number
K083562
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: VISUAL INSPECTION, FUNCTIONAL INSPECTION, AND DEVICE HISTORY REVIEW. RESULTS: VISUAL INSPECTION: NO DEVICE WAS RETURNED (PLATE AND SCREWS REMOVED AND NOT RETURNED) SO NO INSPECTION COULD BE PERFORMED. DEVICE HISTORY REVIEW: NO MANUFACTURING FILES WERE REVIEWED BECAUSE NO LOT NUMBER WAS PROVIDED. CONCLUSION: A FOUR LEVEL AVIATOR PLATE WAS REPORTED TO HAVE A MAJORITY OF THE SCREWS PULLED OUT OF THE VERTEBRAL BODY, WHILE STILL LOCKED IN THE PLATE (BASED ON SALES REP CONVERSATION WITH THE SURGEON). A REVISION SURGERY WAS PERFORMED THAT REPLACED THE PLATE AND SCREWS. THE PLATE WAS NEVER RETURNED. THERE WAS NO INFORMATION AVAILABLE EXCEPT THE SURGEON'S NAME AND THE PATIENT'S GENDER. NO INFORMATION ON THE PATIENT'S LIFESTYLE WAS MADE AVAILABLE AND NO X-RAYS AND/OR PRODUCT CONFIRMING THE FAILURE WERE AVAILABLE. MULTIPLE ATTEMPTS TO COLLECT ADDITIONAL INFORMATION WERE MADE WITH NO SUCCESS.

Additional Manufacturer Narrative · 1

NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT, "FAILED FOUR LEVEL ANTERIOR CERVICAL FUSION. ORIGINAL DATE OF SURGERY WAS IN (B)(6) 2013. DR PREFORMED A FOUR LEVEL ANTERIOR CERVICAL FUSION UTILIZING THE AVIATOR PLATING SYSTEM AND C-PLUG CERVICAL GRAFT. PATIENT PRESENTED 4-6 WEEKS POST OP WITH THE MAJORITY OF THE PLATE SEPARATED FROM THE VERTEBRAL BODIES. ALL SCREWS APPEARED LOCKED TO THE PLATE. ALMOST THE ENTIRE CONSTRUCT PULLED OUT. THE REVISION SURGERY WAS ON (B)(6). THE PLATE WAS REMOVED AND THE SITE WAS RE-INSTRUMENTED UTILIZING A NEW AVIATOR PLATE AND LARGE DIAMETER SCREWS. THE PATIENT WAS ALSO FIXED POSTERIORLY UTILIZING THE OASYS SYSTEM."

Description of Event or Problem · 1

IT WAS REPORTED THAT, "FAILED FOUR LEVEL ANTERIOR CERVICAL FUSION. ORIGINAL DATE OF SURGERY WAS IN (B)(6) 2013. DR. PREFORMED A FOUR LEVEL ANTERIOR CERVICAL FUSION UTILIZING THE AVIATOR PLATING SYSTEM AND C-PLUG CERVICAL GRAFT. PATIENT PRESENTED 4-6 WEEKS POST OP WITH THE MAJORITY OF THE PLATE SEPARATED FROM THE VERTEBRAL BODIES. ALL SCREWS APPEARED LOCKED TO THE PLATE. ALMOST THE ENTIRE CONSTRUCT PULLED OUT. THE REVISION SURGERY WAS ON (B)(6). THE PLATE WAS REMOVED AND THE SITE WAS RE-INSTRUMENTED UTILIZING A NEW AVIATOR PLATE AND LARGE DIAMETER SCREWS. THE PATIENT WAS ALSO FIXED POSTERIORLY UTILIZING THE OASYS SYSTEM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105153 AVIATOR ASSY FOUR LEVEL PLATE SIZE 56 ANTERIOR CERVICAL PLATE KWQ STRYKER SPINE-FRANCE UNK

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R