FDA Adverse Event Injury Summary report: N

SIMPLEX TOBRAMYCIN 1 PK

MDR report key: 3002694 · Received March 13, 2013

Report

Report Number
0002249697-2013-00974
Event Type
Injury
Date Received
March 13, 2013
Date of Event
February 3, 2013
Report Date
February 22, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LOD
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT. THE SUBJECT DEVICE IS NOT CLEARED FOR SALE IN THE U.S., BUT A SIMILAR DEVICE IS COMMERCIALLY AVAILABLE IN THE U.S.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORDS INDICATES THAT THE REPORTED DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE COMPLAINT HISTORY REVIEW INDICATED THAT THERE WERE NO SIMILAR EVENTS FOR THE REPORTED LOT. THE EVENT COULD NOT BE CONFIRMED. VISUAL INSPECTION AND FUNCTIONAL TESTING WAS COMPLETED ON THE THREE RETAIN SAMPLES OF THIS LOT. THE RESULTS WERE SATISFACTORY AND WITHIN SPECIFICATION. THE MIXING CHARACTERISTICS AND WORKING PROPERTIES OF SURGICAL SIMPLEX BONE CEMENTS ARE INFLUENCED PRIMARILY BY THE TEMPERATURE OF THE LIQUID AND POWDER COMPONENTS AT THE TIME OF MIXING AND BY THE TEMPERATURES OF THE UTENSILS WITH WHICH IT CONTACTS DURING MIXING E.G. MIXING BOWLS, CEMENT INTRODUCERS ETC. GENERALLY, HIGHER TEMPERATURES ACCELERATE THE POLYMERISATION REACTION AND LOWER TEMPERATURES DELAY IT. OTHER FACTORS WHICH CAN AFFECT SETTING TIME ARE MIXING TECHNIQUE (SPEED, USE OF VACUUM, CENTRIFUGATION), THOROUGHNESS OF MIXING, COMPLETE UTILIZATION OF ALL OF THE POWDER & LIQUID AND CARE TO AVOID INCLUSION OF ANY EXTRANEOUS MATERIAL SUCH AS BLOOD OR STERILISATION SOLUTIONS INTO THE MIX. MIXING PROCESS/TECHNIQUE ISSUES ARE HIGHLIGHTED IN THE OR HANDBOOK. THE MIXING PROPERTIES OF THE RETAIN SAMPLES OF THE REPORTED LOT CODE TJT056 WERE TESTED AND SHOW THAT ALL REQUIRED SPECIFICATIONS ARE MET. IT WAS NOT POSSIBLE TO REPLICATE THIS EVENT. NO FURTHER INVESTIGATION IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CEMENT HARDENED PREMATURELY (WITHIN 6 MINUTES) WHILE PLACING THE IMPLANT STEM OF THE HIP REPLACEMENT BEFORE THE STEM WAS COMPLETELY DOWN THE FEMUR SHAFT. AS A RESULT THE HIP REPAIR BECAME A REVISION CAUSING A HIP FRACTURE AND SURGICAL DELAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CEMENT HARDENED PREMATURELY (WITHIN 6 MINUTES) WHILE PLACING THE IMPLANT STEM OF THE HIP REPLACEMENT BEFORE THE STEM WAS COMPLETELY DOWN THE FEMUR SHAFT. AS A RESULT THE HIP REPAIR BECAME A REVISION CAUSING A HIP FRACTURE AND SURGICAL DELAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105463 SIMPLEX TOBRAMYCIN 1 PK IMPLANT LOD STRYKER ORTHOPAEDICS-MAHWAH TJT056

Patients

Seq Age Sex Outcome Treatment
1 Other