FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 3002581 · Received February 14, 2013

Report

Report Number
1824206-2013-00988
Event Type
Malfunction
Date Received
February 14, 2013
Date of Event
January 28, 2013
Report Date
January 28, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECH FOUND THE GROUND PIN WAS BENT AND LOOSE ON THE POWER CORD. HE REPLACED THE POWER CORD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THE BED HAD A LOSS OF GROUND CONTINUITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65106 TOTALCARE BED AC POWERED ADJUSTABLE HOSP BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1