FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3002565 · Received February 15, 2013

Report

Report Number
3008642652-2013-00385
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 28, 2013
Report Date
February 8, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK THERAPY ELECTRODE MESSAGES) WAS CONFIRMED. UPON INVESTIGATION THE PULSE WIRE IN THE CABLE CONNECTING THE DISTRIBUTION NODE (DN) AND ECG "B" WAS OPEN, WHICH CAUSED THE REPORTED ALARMS. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE ON THE CABLE SECTION. NO ADVERSE EVENT RESULTED FROM THE OPEN PULSE WIRE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

THE DAUGHTER OF A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT WAS RECEIVING CHECK THERAPY ELECTRODE MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
66513 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR