FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3002521 · Received February 15, 2013

Report

Report Number
3008642652-2013-00399
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 4, 2013
Report Date
February 11, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CONNECTOR) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT'S TRUNK CABLE CONNECTOR WAS BROKEN AND THE LOCKING NUT WAS CRACKED. AS A RESULT, THE ELECTRODE BELT WAS UNABLE TO STAY SECURELY CONNECTED TO A MONITOR. THE ROOT CAUSE FOR THE DAMAGED CONNECTOR COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A ZOLL TERRITORY MANAGER CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) FEMALE PATIENT'S ELECTRODE BELT HAD A DAMAGED CONNECTOR. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67570 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR