FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 3002521
·
Received February 15, 2013
Report
- Report Number
- 3008642652-2013-00399
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 4, 2013
- Report Date
- February 11, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CONNECTOR) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT'S TRUNK CABLE CONNECTOR WAS BROKEN AND THE LOCKING NUT WAS CRACKED. AS A RESULT, THE ELECTRODE BELT WAS UNABLE TO STAY SECURELY CONNECTED TO A MONITOR. THE ROOT CAUSE FOR THE DAMAGED CONNECTOR COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A ZOLL TERRITORY MANAGER CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) FEMALE PATIENT'S ELECTRODE BELT HAD A DAMAGED CONNECTOR. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67570 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |