FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3002519 · Received February 15, 2013

Report

Report Number
3008642652-2013-00365
Event Type
Malfunction
Date Received
February 15, 2013
Date of Event
January 22, 2013
Report Date
February 6, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY CHARGER PROBLEM) WAS CONFIRMED. UPON EVALUATION, THE POWER BRICK WAS DEFECTIVE. THE CAUSE OF THE DEFECTIVE POWER BRICK IS A DAMAGED CONNECTOR. THE CONNECTOR PINS WERE RECESSED INTO THE CONNECTOR. THE ROOT CAUSE OF THE RECESSED PINS CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT CHARGER/MODEM.

Description of Event or Problem · 1

A PHYSICIAN ASSISTANT (PA) CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE'S BATTERY CHARGER MODEM WOULD NOT STAY POWERED ON. THE PA REPORTED THAT THE CHARGER HAD A LOOSE POWER CONNECTION. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
67753 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 58 YR