LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-00365
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 22, 2013
- Report Date
- February 6, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY CHARGER PROBLEM) WAS CONFIRMED. UPON EVALUATION, THE POWER BRICK WAS DEFECTIVE. THE CAUSE OF THE DEFECTIVE POWER BRICK IS A DAMAGED CONNECTOR. THE CONNECTOR PINS WERE RECESSED INTO THE CONNECTOR. THE ROOT CAUSE OF THE RECESSED PINS CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT CHARGER/MODEM.
A PHYSICIAN ASSISTANT (PA) CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) MALE'S BATTERY CHARGER MODEM WOULD NOT STAY POWERED ON. THE PA REPORTED THAT THE CHARGER HAD A LOOSE POWER CONNECTION. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 67753 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |