LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-00409
- Event Type
- Malfunction
- Date Received
- February 15, 2013
- Date of Event
- January 29, 2013
- Report Date
- February 11, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (ADD GEL REPLACE BELT MESSAGES) WAS CONFIRMED. UPON INVESTIGATION THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THIS RESULTED IN AN INTERMITTENT CONNECTION BETWEEN THE CABLE AND ON. THE ROOT CAUSE OF THE DAMAGED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A (B)(6) YEAR OLD MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT SEVERAL ADD GEL MESSAGES. HOWEVER, HE REPORTS GEL WAS NEVER DEPLOYED. ZOLL CUSTOMER SUPPORT PROVIDED THE PATIENT WITH A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 66255 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |