INFUSOR
Report
- Report Number
- 1416980-2013-06000
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- February 25, 2013
- Report Date
- February 25, 2013
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE CONTAINING NO FLUID IN THE RESERVOIR. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGN OF PHYSICAL ABNORMALITY. AN ACCURACY FLOW TEST WAS PERFORMED ON THE SAMPLE. THE INFUSOR PRODUCED FUNCTIONAL RESULTS WITHIN THE SPECIFICATION RANGE; THE DEVICE PERFORMED AS EXPECTED. THE REPORTED CONDITION OF AN OVERINFUSION WAS NOT CONFIRMED. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW-UP WILL BE SUBMITTED.
IT WAS REPORTED TO BAXTER (B)(4) THAT A 2, 3, 5 ML/HR MULTIRATE INFUSOR OVERINFUSED DURING PATIENT USE. THE TOTAL FILL VOLUME OF 100 ML WAS INFUSED OVER THE COURSE OF 20 HOURS WHEN SET ON 3 ML/HR FLOW RATE. THE DEVICE WAS FILLED WITH A SOLUTION OF ROPIVACAINE HYDROCHLORIDE HYDRATE. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 105002 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 12M068 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ROPIVACAINE HYDROCHLORIDE HYDRATE |