FDA Adverse Event Malfunction Summary report: N

STERLING¿

MDR report key: 3002383 · Received March 13, 2013

Report

Report Number
2134265-2013-01666
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 22, 2013
Report Date
February 25, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
DQY
PMA / PMN Number
K053116
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER: THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE A BALLOON RUPTURE OCCURRED. THE CALCIFIED TARGET LESION WAS LOCATED IN THE SHUNT. A 6.0MMX40MMX40CM (4F) STERLING BALLOON CATHETER WAS SELECTED AND ADVANCED TO TREAT THE TARGET LESION. THE BALLOON RUPTURED AT 6 ATMS ON THE SECOND INFLATION. THE ATMS OF THE FIRST INFLATION ARE UNKNOWN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
105668 STERLING¿ CATHETER, PERCUTANEOUS DQY BOSTON SCIENTIFIC - MAPLE GROVE H74939032604040 15637968

Patients

Seq Age Sex Outcome Treatment
1